6 3.2 Panel Mount 1. Prepare panel cut-out of 92.0 mm by 92.0 mm2. Remove back cover of transmitter and slide it through panel cut-outPane

Strany 3 - Preface

7 4 DISPL AY AND KEYPAD F UNCTIONS 4.1 Display The LCD has a primary and secondary display.  The primary display shows the measured pH, mV or R

Strany 4 - TABLE OF CONTENTS

8 4.2 Keypad The four-button keypad allows easy and quick operations of the Transmitter. Key Function CAL Brings you directly into the Calibration

Strany 5 - 1 INTRODUCTION

9 5 CALIBRATION 5.1 Preparing the Transmitter for Calibration Before starting calibration, make sure you are in the correct measurement

Strany 6 - 2 PREPARATION

10 5.2 pH Calibration A complete 2-point CAL is required for this Transmitter. If Calibration is aborted after 1st calibration point, transmitter

Strany 7 - Position)

11 5.3 Relative mV Calibration 1. From the ORP MEAS mode, press CAL key. CAL indicator appears at top of LCD. Primary display show

Strany 8

12 SETUPHOLDSETUPHOLDpH READYpH 6 ADVANCED SETUP FUNCTIONS The advanced setup mode lets you customize your transmitter‟s preferences and de

Strany 9 - 3 INSTALL ATION

13 SETUPHOLD6.2 Output Range Setting Output of pH 500 transmitter is via the 2-wire power supply loop. Both the 4 mA and 20 mA ou

Strany 10 - 3.2 Panel Mount

14 6.3 Temperature Compensation Setting (Unavailable in ORP mode) pH values other than pH 7.00 are affected by temperature. Under varyi

Strany 11 - 4.1 Display

15 2. Press ENT key. Primary display shows „ATC‟ and secondary display shows „On‟ or „OFF‟. Use ▲ or ▼ keys to select „ATC Off‟. Press ENT

Strany 12 - 4.2 Keypad

AQUAfast, Cahn, ionplus, KNIpHE, No Cal, ORION, perpHect, PerpHecT, PerpHecTion, pHISA, pHuture, Pure Water, Sage, Sensing the Future, SensorLink, R

Strany 13 - 5 CALIBRATION

16 SETUPHOLDSETUPHOLDSETUPHOLD°CATCSETUPHOLDSETUPHOLDSETUPHOLD Temperature Compensation Setting Chart – alpha pH 500 Transmitter

Strany 14 - 5.2 pH Calibration

17 SETUPHOLDSETUPHOLDSETUPHOLD6.4 HOLD Current Setting When Transmitter is in CAL or SETUP modes, it automatically goes into a „HOLD‟ mode.

Strany 15

18 6.6 Configuration 6.6.1 Selecting pH or ORP Mode of Operation The alpha pH 500 transmitter is not only a pH transmitter, but also an ORP tran

Strany 16 - 6 ADVANCED SETUP FUNCTIONS

19 3. LCD displays „ASY‟ (asymmetrical operation). Use ▲ or ▼ keys to select for „SYS‟ (symmetrical mode of operation). Press ENT key again. 4. Pr

Strany 17 - 6.2 Output Range Setting

20 7 PROBE CARE AND M AINT EN AN CE Since your pH electrode is susceptible to dirt and contamination, clean it every one to three months dependin

Strany 18 - (Unavailable in ORP mode)

21 Reactivating the pH electrode If stored and cleaned properly, your pH electrode should be ready for immediate use. However, a dehydra

Strany 19

22 8 TROUBLE SH OOTING GUIDE Problem Cause Solution Power on but no display a). Loose connections b). Cables not in correct polarity (+ and – posit

Strany 20

23 9 SPECIFICATIONS SPECIFICATIONS DESCRIPTIONS pH Range 0.00 to 14.00 pH Resolution / Accuracy 0.01 pH /  0.01 pH Temperature Range -10.0 to

Strany 21 - 6.4 HOLD Current Setting

24 10 TECHNIC AL DIM ENSIONS

Strany 22

25 11 ACCESSORIES THERMO SCIENTIFIC Replacement Transmitter and Transmitter accessories Ordering Code No. Item TSPHCTP0500 Alpha pH/ORP 2-wire LC

Strany 23

Preface This manual serves to explain the use of the Alpha pH 500 transmitter. It functions in two ways, firstly as a step by step guide to help you

Strany 24

26 EUTECH INSTRUMENTS Replacement Transmitter and Transmitter accessories Ordering Code No. Item 56717-20 pH 500 pH/ORP 2-wire LCD Transmitter wit

Strany 25

27 12 WARR ANTY This transmitter is supplied with a one-year warranty against significant deviations in material and workmanship from date of

Strany 26 - 8 TROUBLE SH OOTING GUIDE

Water Analysis InstrumentsNorth America166 Cummings CenterBeverly, MA 01915 USAToll Free: 1-800-225-1480Tel: 1-978-232-6000Dom. Fax: 1-978-232-6015Int

Strany 27 - 9 SPECIFICATIONS

TABLE OF CONTENTS 1 INTRODUCTION 1 2 PREPARATION 2 2.1 Power Supply Requirements (SL2 Position) 3 2.2 Connecting the Electrode & Temper

Strany 28 - 10 TECHNIC AL DIM ENSIONS

1 1 INTRODUCTION Thank you for selecting an Alpha pH 500 pH/ORP Transmitter. This transmitter is a microprocessor-based instrument that is designe

Strany 29 - 11 ACCESSORIES

2 2 PREPARATION Remove screws from the four corners at the back of the Transmitter, and remove back cover. Connectors should be exposed as follows

Strany 30 - Calibration Solutions

3 2.1 Power Supply Requirements (SL2 Position) This transmitter requires a 12 to 24V DC power supply. Other Transmitters and/or a chart recorder ma

Strany 31 - 13 RETURN OF ITEMS

4 2.2.2 To connect temperature probe: For Automatic Temperature Compensated (ATC) pH readings, a 100 Pt RTD temperature probes can be

Strany 32 - Water Analysis Instruments

5 3 INSTALL ATION 3.1 Wall Mount