TechnologyMadeEasy...Instruction Manual αlpha CON 2000 Conductivity/ Total Dissolved Solids Controller / Transmitter 68X216843 Rev 1 01/
Instruction Manual αlpha CON 2000 42.2 Measurement and control system A typical measurement system consists of: • A conductivity process trans
Instruction Manual αlpha CON 2000 52.3 Unit overview Wall mounting version Panel mounting version
Instruction Manual αlpha CON 2000 62.3.1 Display Overview The LC display shows two alpha-numerical fields for parameters and measured values as
Instruction Manual αlpha CON 2000 72.3.3 LED indicators Relay indicators If REL key is pressed the LED (A, B or W) indicates to which Rela
Instruction Manual αlpha CON 2000 82.3.5 Menu overview CCD “000” = Check calibration parameters (View only mode) CCD “11” = Calibration mode SC
Instruction Manual αlpha CON 2000 93 ASSEMBLY AND INSTALLATION 3.1 Mounting the unit Wall mounting version approx. 14 [.55]39 [1.54]Hole
Instruction Manual αlpha CON 2000 10Panel mounting version Flat gasket 1mm [.04](to be inserted by customer)Panel cut outUNIT: MM [INCH] Trans
Instruction Manual αlpha CON 2000 113.2 Connection Diagram Caution: Ensure electrical mains are disconnected before proceeding. Connections fo
Instruction Manual αlpha CON 2000 12 NOTE: a) Switch or circuit breaker shall included in the building installation. b) It shall be in clos
Instruction Manual αlpha CON 2000 13Connections for panel mounting version 1. AC mains live wire **17. 4 Cell / 2 Cell type Conductivity Inp
Instruction Manual αlpha CON 2000 14 NOTE: 1) Switch or circuit breaker shall included in the building installation. 2) It shall be in close
Instruction Manual αlpha CON 2000 154 NORMAL OPERATION 4.1 Measurement mode When the transmitter is powered on, the display first shows all
Instruction Manual αlpha CON 2000 165 CALIBRATION MODE You can access the Calibration mode directly from the Measurement mode by pressing the
Instruction Manual αlpha CON 2000 175.2 Calibration This transmitter features a one-point calibration. Note: The calibration is always carrie
Instruction Manual αlpha CON 2000 18 Note: If you entered the Calibration mode from the Setup mode, the transmitter will return to the setup menu
Instruction Manual αlpha CON 2000 19MEASSETUPSETUPSETUPSETUPSETUPSETUPSETUPSETUPSETUPSETUPHOLDHOLDHOLDHOLDHOLDHOLDHOLDHOLDHOLDHOLDµS444444444444°
Instruction Manual αlpha CON 2000 206.2 Temperature compensation (TC) sub-function This sub-function allows you to select the correct temperatu
Instruction Manual αlpha CON 2000 216.3 Setting temperature (Set °C°F) sub-function SETUPSETUPSETUPHOLDHOLDHOLD444ENTSETUPSETUPHOLDHOLD44ENT°C
Instruction Manual αlpha CON 2000 226.4 Control Relay A / Relay B (SP1/SP2) sub-function The SP1 sub-function determines the operating paramete
Instruction Manual αlpha CON 2000 23 5. Setting the on-delay time lag: press the ▲ or ▼ key to enter the on-delay time for set point 1 (set poin
Preface This manual serves to explain the use of the αlpha CON 2000 Series. The manual functions in two ways, firstly as a step by step guide to help
Instruction Manual αlpha CON 2000 246.5 Controller (CNTR) sub-function The CNTR sub-function determines the controller’s parameters. SETUPSET
Instruction Manual αlpha CON 2000 251. Select the “CNTR” subfunction, then press the ENT key. 2. Selecting the controller type: press the ▲ or
Instruction Manual αlpha CON 2000 26 − If the proportional controller type is set to pulse frequency control (PFC): 9. Selecting the relay sta
Instruction Manual αlpha CON 2000 276.6 Current Output 1 sub-function In this subfunction you set the current output range of the transmitter f
Instruction Manual αlpha CON 2000 286.7 Current Output 2 sub-function In this sub-function you set the current output range of the transmitter
Instruction Manual αlpha CON 2000 296.8 Wash relay (WASH) sub-function In this sub-function you set the parameters for the wash relay. SETUP
Instruction Manual αlpha CON 2000 30SETUPSETUPSETUPSETUPSETUPSETUPSETUPSETUPHOLDHOLDHOLDHOLDHOLDHOLDHOLDHOLDµSµSµSµSmSmSmS41234567ENTENTENTENTENT
Instruction Manual αlpha CON 2000 316.10 Configuration (CNFG) sub-function In this sub-function you configure the transmitter to suit your req
Instruction Manual αlpha CON 2000 32 6. Selecting the alarm contact type: press the ▲ or ▼ key to select the alarm contact type: − “STDY” = st
Instruction Manual αlpha CON 2000 337 RELAY MODES You can control devices connected to Relay A, Relay B or wash relay via the front panel of th
Safety Information This Eutech Instruments Controller/ Transmitter shall be installed and operated only in the manner specified in the Instruction m
Instruction Manual αlpha CON 2000 347.2 Manual relay mode In manual relay mode, you can manually turn “on” and “off” the control devices connec
Instruction Manual αlpha CON 2000 358 TECHNICAL SPECIFICATIONS 8.1 General specifications Conductivity Measuring range No Measuring range
Instruction Manual αlpha CON 2000 36Alarm Functions Function (switchable) Latching / pulse Pickup delay 0 to 2000 seconds Switching voltage M
Instruction Manual αlpha CON 2000 37 Mechanical Specifications Dimensions 144 x 144 x 111.5 mm (5.67 x 5.67 x 4.39 inch) Weight 950 g Material
Instruction Manual αlpha CON 2000 389 ACCESSORIES 9.1 Replacement Unit Product Description Eutech Instruments Order Code αlpha CON 2000 Contr
Instruction Manual αlpha CON 2000 3910 GENERAL INFORMATION 10.1 Warranty Eutech Instruments supplies this product with a warranty of one year
Instruction Manual αlpha CON 2000 4010.5 Maintenance and Cleaning Maintenance The αlpha CON 2000 contains no user repairable components. P
Instruction Manual αlpha CON 2000 4111 APPENDICES 11.1 Appendix 1 – Unit fuse and jumper settings Caution! Before opening the unit to replac
Instruction Manual αlpha CON 2000 42Panel mounting version (view from top)
Instruction Manual αlpha CON 2000 4311.2 Appendix 2 – Conductivity of various aqueous solutions at 25 °C / 77 °F Conductivity Resistivity Ult
TABLE OF CONTENTS 1 INTRODUCTION...
Instruction Manual αlpha CON 2000 4411.3 Appendix 3 – Simple explanation on the function of hysteresis 100 120 1880 1900RELAY OFFHYSTERESIS BA
Instruction Manual αlpha CON 2000 4511.4 Appendix 4 – General instructions concerning Controller Setting 11.4.1 Control characteristic of Cont
Instruction Manual αlpha CON 2000 46 11.4.4 Control signal of Pulse length Controllers RelayONtONtOFFPulse Length TOFFTime [s] The output r
Instruction Manual αlpha CON 2000 47 11.4.5 Control signal of Pulse Frequency Controllers RelayONtONtOFFPulse Length TOFFTime [s] The outpu
Instruction Manual αlpha CON 2000 4811.5 Appendix 5 – Abbreviations used in menu displays Abbreviation Meaning Abbreviation Meaning MEAS Meas
For more information on Eutech Instruments’ products, contact your nearest distributor or visit our website listed below:
10.2 PACKAGING / SCOPE OF DELIVERY... 39 10.3 RET
Instruction Manual αlpha CON 2000 11 INTRODUCTION 1.1 At the very beginning We thank you for having purchased the Eutech Instruments αlpha CO
Instruction Manual αlpha CON 2000 21.3 Safety intsructions The αlpha CON 2000 should be installed and operated only by personnel familiar with
Instruction Manual αlpha CON 2000 32 PRODUCT DESCRIPTION 2.1 Description of unit The Eutech Instruments αlpha CON 2000 is used for measuring