Instruction Manual CyberScan DO 1500 Dissolved Oxygen Meter 68X292336 Rev. 1 01/04 TechnologyMad

Strany 2 - PREFACE

Instruction Manual CyberScan DO 1500 7 P 1.0: View calibration data- View the last calibrated cal factor and 0% offs

Strany 3

Instruction Manual CyberScan DO 1500 8 4.1.1. Setup P 1.0: View the Calibration Data This setup menu allows you to view

Strany 4 - 1. INTRODUCTION

Instruction Manual CyberScan DO 1500 9 2. Press enter to access View Last calibrated Cal factor setup page.  press to

Strany 5 - 2.1. Connectors

Instruction Manual CyberScan DO 1500 10 3. If the unit has not been standardised, a series of dashes will appear on the

Strany 6 - 3.1. The Electrode

Instruction Manual CyberScan DO 1500 11 4.1.2. Setup P 2.0: Set Pressure The setup option allows you to set the atmo

Strany 7 - 3.2. Display/ Keys

Instruction Manual CyberScan DO 1500 12 3. Press enter to go to the ‘Pressure adjustment’ setup page as shown below. 4

Strany 8 - 3.3. Screen Display

Instruction Manual CyberScan DO 1500 13 4.1.3. Setup P3.0: Set the Salinity Value The setup option allows you to set th

Strany 9 - 4. SETUP PROCEDURE CHECK

Instruction Manual CyberScan DO 1500 14 3. Press enter to go to the ‘Salinity adjustment’ setup page as shown below. 4

Strany 10

Instruction Manual CyberScan DO 1500 15 4.1.4. Setup P4.0: Select Number of User Cal Points for %Sat calibration

Strany 11

Instruction Manual CyberScan DO 1500 16 3. Press enter to accept the current value or press  or  to adjust and set

Strany 12

PREFACE Thank you for selecting the Eutech Instruments CyberScan DO 1500 bench meter. The instruction manual serves to explain the use of the Cybe

Strany 13

Instruction Manual CyberScan DO 1500 17 4.1.5. Setup P5.0: Enable/ Disable Auto Cal This setup option allows you to e

Strany 14

Instruction Manual CyberScan DO 1500 18 3. Press  or  to select YES or NO. 4. Press enter key to confirm selection a

Strany 15

Instruction Manual CyberScan DO 1500 19 4.1.6. Setup P6.0: Select the Temperature Unit sel ect  press to set valu

Strany 16 - Salinity

Instruction Manual CyberScan DO 1500 20 3. Press  or  key to select °C or °F. 4. Press enter key to confirm selectio

Strany 17

Instruction Manual CyberScan DO 1500 21 4.1.7. Setup P7.0: Enable/ Disable Stability Indicator This setup option allow

Strany 18

Instruction Manual CyberScan DO 1500 22 3. Press  or  key to select ON or OFF. 4. Press enter key to confirm selecti

Strany 19

Instruction Manual CyberScan DO 1500 23 4.1.8. Setup P8.0: Baud Rate This setup option allows you to set the baud rate

Strany 20

Instruction Manual CyberScan DO 1500 24 3. Press  or  key to toggle between 4800, 9600, 19200 or 38400. 4. Press

Strany 21

Instruction Manual CyberScan DO 1500 25 4.1.9. Setup P9.0: Parity Bit This setup option allows you to set the parity bi

Strany 22

Instruction Manual CyberScan DO 1500 26 3. Press  or  key to toggle between toggle between 0 (none), 1 (odd) or 2 (eve

Strany 23

TABLE OF CONTENTS 1. INTRODUCTION 1 2. GETTING STARTED 2 2.1. Connectors 2 3. USING THE METER 3 3.1. The Electrode 3 3.2. Display/ Keys

Strany 24

Instruction Manual CyberScan DO 1500 27 4.1.10. Setup P10.0: Stop Bit This setup option allows you to set the stop bit

Strany 25

Instruction Manual CyberScan DO 1500 28 3. Press  or  key to toggle between 1 or 2. 4. Press enter key to confirm se

Strany 26

Instruction Manual CyberScan DO 1500 29 4.1.11. Setup P11.0: Print Data This setup option allows you to print current d

Strany 27

Instruction Manual CyberScan DO 1500 30 3. Press  or  key to toggle between Curr or Store. 4. Press enter key to con

Strany 28 - pr ess to set val ue

Instruction Manual CyberScan DO 1500 31 4.1.12. P12.0: Clear Stored Data This setup option allows you to clear all stor

Strany 29

Instruction Manual CyberScan DO 1500 32 3. Press  or  key to toggle between YES or NO. 4. Press enter key to confirm

Strany 30

Instruction Manual CyberScan DO 1500 33 4.1.13. P13.0: Clear Buffer Values This setup option allows you to clear the st

Strany 31

Instruction Manual CyberScan DO 1500 34 3. Press  or  key to toggle between YES or NO. 4. Press enter key to confirm

Strany 32

Instruction Manual CyberScan DO 1500 35 5. USER CALIBRATION 5.1. %Sat – One Point calibration (Auto) User can do the

Strany 33

Instruction Manual CyberScan DO 1500 36 5.2. %Sat – Two Point calibration (Auto) User must do the 0.0%Sat cal first. T

Strany 34

Instruction Manual CyberScan DO1500 11. INTRODUCTION Thank you for selecting a Eutech Instruments CyberScan Dissolved Ox

Strany 35

Instruction Manual CyberScan DO 1500 37 5.3. %Sat – One Point calibration (Manual) User can calibrate to the know valu

Strany 36

Instruction Manual CyberScan DO 1500 38 5.4. mg/L– One Point calibration (Auto) User can do calibration in mg/L. 1.

Strany 37

Instruction Manual CyberScan DO 1500 39 5.5. mg/L– One Point calibration (Manual) User can do calibration in mg/L. 1

Strany 38 - Standardize

Instruction Manual CyberScan DO 1500 40 5.6. ATC Calibration User can adjust the ATC temperature offset. 1. From th

Strany 39 - Standar dize

Instruction Manual CyberScan DO 1500 41 saveprintentersaveview6. MEMORY 6.1. Store Value into Memory In any measur

Strany 40

Instruction Manual CyberScan DO 1500 42 stdprintenter7. PRINT DATA 7.1. Printing Data Depending on the print option i

Strany 41

Instruction Manual CyberScan DO 1500 43 8. DO THEORY The polarographic type Oxygen probe used with the CyberScan DO 1500

Strany 42

Instruction Manual CyberScan DO 1500 44 9. METER SPECIFICATIONS Description CyberScan DO 1500 DO range 0.00 to 60.

Strany 43

Instruction Manual CyberScan DO 1500 45 10. CLEANING This meter requires no regular maintenance, but it is recommended t

Strany 44 - 6. MEMORY

Instruction Manual CyberScan DO 1500 46 12. ACCESSORIES Consult your Authorised Distributors for these items. 12.1.

Strany 45 - 7. PRINT DATA

Instruction Manual CyberScan DO 1500 2 2. GETTING STARTED 2.1. Connectors 1. Review the layout and arrangement of th

Strany 46 - 8. DO THEORY

Instruction Manual CyberScan DO 1500 47 13. WARRANTY Eutech Instruments supplies this bench meter with a 3-year warranty

Strany 47 - 9. METER SPECIFICATIONS

Instruction Manual CyberScan DO 1500 48 14. NOTICE OF COMPLIANCE Warning This meter generates, uses, and can radiate r

Strany 48 - 11. TROUBLESHOOTING

For more information on Eutech Instruments products, contact your nearest Eutech Instruments distributor or visit

Strany 49 - 12. ACCESSORIES

Instruction Manual CyberScan DO 1500 3 3. USING THE METER 3.1. The Electrode This meter comes equipped with a self

Strany 50 - 13. WARRANTY

Instruction Manual CyberScan DO 1500 4  press to select optionssetup to standardizepress to store/recall datapres

Strany 51 - 14. NOTICE OF COMPLIANCE

Instruction Manual CyberScan DO 1500 5 3.3. Screen Display Familiarise yourself with the layout of the digital screen d

Strany 52

Instruction Manual CyberScan DO 1500 6 stdprintentersaveview4. SETUP PROCEDURE CHECK 4.1. CyberScan DO 1500 Setup Pa