Additional dissolved oxygen calibration notes:
Keep the membrane free from contact with solid objects.
Use with aqueous solutions only.
Do not submerse the probe past the immersion limit (see below).
See Section 22—DO Electrode Maintenance for additional information.
Calibration adjustment is limited to ±40% of the factory default value to
prevent erroneous calibrations.
If % calibration is attempted with a value of 10.1% to 49.9%, a
“Calibration Error” will result. The following table lists corresponding
calibration values in % saturation calibration mode:
% Saturation
(of factory default value)
Calibration Value
less than 10% 0%
10.1% to 49.9% Calibration Error
50% to 150% 100%
Calibration of the concentration mode will only replace the previous
concentration calibration and does not affect the % saturation calibration.
To offset your % satu
ration reading to match another instrument, see
Section 15—Offset Calibration.
Perform daily calibration for best results. New calibration values will
automatically override the previous calibration.
36