6.10 Reset Function
The reset function lets you choose to either reset the
transmitter’s conductivity calibration only or conductivity
calibration plus all setup functions that you might have
changed back to factory default settings. However, the
temperature compensation setting, as in Section 6.2, will
remain unchanged.
From the measurement mode,
1. Press the ENT key to enter the setup mode. Use
the or keys to scroll till LCD displays ‘rSt’ in
the upper display.
2. Press the ENT key to enter the reset page. Use
the or keys to toggle between ‘CAL’ and
‘FCT’ for reset type selection.
CAL – reset conductivity calibration only
FCT – reset conductivity calibration and all
setup functions back to factory default
3. Press the ENT key to confirm your selection.
The lower display will blink momentarily before returning to the reset
function setup mode.
4. Press S and T keys together, to return to the measurement mode.
SETUP
SETUP
SETUP
HOLD
HOLD
HOLD
CAL
FC