5.8 Viewing Electrode Properties
Each time you calibrate the conductivity sensor, the controller re-calculates the
calibration factor of the sensor and shown in the LCD at the end of calibration. The
setup mode allows you to view the last saved calibration factor and cell constant at
any time.
SETUP
HOLD
2
SETUP
ENT
1
ENT
HOLD
SETUP
HOLD
ENT
11
From measurement mode press ENT key to enter setup mode as
described in section 5.1. The LCD shows the first screen of setup mode
(rAng). Press or key to select viewing electrode properties
screen (CdAt).
Press ENT key to view electrode properties (CdAt).
2
Cell constant: The lower display shows ’CEL’ and the upper display
shows the last configured cell constant.
Press ENT key.
3
Calibration factor: The lower display shows ‘FCt’ and the upper
display shows the calibration factor based on the last conductivity
calibration.
NOTE: The upper display shows ‘- - - ‘ if the controller has not been
calibrated yet or the controller has been reset to factory defaults.
Press ENT key to quit from electrode properties screen. The controller
reverts to CdAt screen.
Press or key to access other setup screens or press and key
simultaneously (escape) to return to measurement mode.