Operating Instructions αlpha-CON1000
68X216802
19
7.4 Controller (Cntr) sub-function
You can set the controller’s parameters in this sub-function.
7.4.1 Entering the Controller sub-function
1) Enter Advanced Set-up mode. Push the ENTER key and scroll to Advanced set-up security code
“22”. Push the ENTER key again.
2) Press the or keys to scroll until the upper display shows “Cntr”.
7.4.2 Choosing the controller type (limit or monitor)
This mode lets you choose your controller type: limit control or control off.
- Use control off to operate controller as a monitor only or to prevent relays from switching.
1) Follow directions in 7.6.1 to enter Controller mode.
2) Press the ENTER key. The upper display shows the current controller type and the lower display
shows “tyP”.
3) Press the or keys to select your controller type.
- L.Ct = limit value pickup (on/off control).
- oFF = controller off.
4) Press the ENTER key to confirm your selection.
5) Proceed to 7.6.3, or return to Measurement mode by pressing the and keys simultaneously
(escape).
CntrCntr
HOLD SETUP
L.CtL.Ct
tyPtyP
HOLD SETUP
HOLD SETUP
dEEndEEn
rELrEL
HOLD SETUP
oFFoFF
tyPtyP