Thermo Scientific Alpha pH 560 33
Figure 5–26. Configuration Settings
11
From pH (or ORP) measurement mode press ENT key to
enter setup mode as described in page 24. The LCD shows the
first screen of setup mode (OFS). Press ▲ or ▼ key to select
configuration settings screen (Cnfg).
Press ENT key to access configuration settings (Cnfg).
2
Enable/disable LCD backlight: The upper display shows
‘LitE’. The lower display shows last configured backlight
selection (‘On’ or ‘OFF’).
Press ▲ or ▼ key to enable (LitE On) or disable (LitE OFF)
backlight of the LCD.
Press ENT key to confirm your selection.
3
Select measurement mode: The upper display shows the last
configured measurement mode (‘PH’ or ‘OrP’).
Press ▲ or ▼ key to select your desired measurement mode.
Press ENT key to confirm your selection.
NOTE: When the measurement mode is changed, the relay
operation parameters (described in page 28) will be reset back