7.6 Configuration (ConF) sub-function
This group of parameters lets you configure the controller to suit your requirements.
7.6.1 Entering the Configuration sub-function
1) Enter Advanced Set-up mode. Press the ENTER key and scroll to Advanced Set-up
security code “22”. Press the ENTER key again.
2) Press the or key to scroll until the upper display shows “ConF”.
7.6.2 Selecting pH, ORP% or ORP mV measurement
This parameter group lets you select pH, mV, or % readings, and input type symmetrical or
asymmetrical. Symmetrical mode is used in conjunction with a potential matching pin – in an
electrically noisy environment.
Please refer to section 3, page 6 for a graphical representation.
1) Follow directions in 7.6.1 to enter Configuration mode.
2) Press the ENTER key. Scroll with the or key until the upper display shows the control
type (pH, ORP mV or ORP %) and the lower display shows symmetrical (SY) or
asymmetrical (ASY) input type. Choose ORP without the mV indicator for ORP% readings.
3) Proceed to 7.6.3, or return to Measurement mode by pressing the and keys
simultaneously (escape).
SETUP
HO LD
SETUP
HO LD
Sy
SETUP
HO LD
SETUP
HO LD
ASy
SETUP
HO LD
Ph
SETUP
HO LD
Sy
SETUP
HO LD
Sy
mV
Ph
Asy ASy