Instruction Manual Salt 6
24
6.5 Restore Factory Default Values
This function allows you to reset all parameters to factory default settings. This
clears all calibration data and any other setup functions you might have
changed.
IMPORTANT: Once activated the meter’s settings and calibration data will be
erased and always exercise caution as meter reset is not reversible.
1) Enter the advanced setup as described
in Section 6.1.
2) Press or key until [UrSt] appears
on the LCD. Press ENTER key.
3) Press or key to select either
[Yes] or [no].
4) Press ENTER key to select.
5) The meter will go back to
measurement mode after the switch-on
initialization as shown in figure of
Section 3.6.
HOLD
ENTER