WATER QUALITY
INSTRUMENTS
14
In-Situ Aqua TROLL 200
Conductivity + Level + Temp
Data Logger
Conductivity measurements can be used to characterize
water quality changes relative to a baseline or to estimate the
concentration of contaminants. Aqua TROLL® instruments simplify
groundwater and surface water monitoring.
Two models available:
• AquaTROLL®100—Measuresandlogsconductivityand
temperature.
• AquaTROLL®200—Measuresandlogsconductivity,waterlevel/
pressure, and temperature. Dynamic density compensation
delivers superior water level accuracy.
Flexible Communications
• Built-inModbus/RS485,SDI-12,and4-20mAsimplifySCADA
and telemetry system integration.
• UsesWin-Situ®softwareplatform.
Extended Deployments
• Titaniumhousingsafeguardsagainstcorrosion.
• Batterieslastaminimumof5yearswhenreadingevery15
minutes—guaranteed.
• TROLL®Shieldantifoulingsystemsignicantlyreduces
biofouling.
Aqua TROLL Specications
Aqua TROLL 200 Aqua TROLL 200
Length
Diameter
Weight
Housing Material
Nose Cone Materiel
Output Options
Fastest Polling Rate
Pressure Range of
Operation
Operating Temperature
Range
Storage Temperature
Power
Internal Battery
External Power
Typical Battery Life
Logging Memory
Log Types
Fastest Reading* Rate
Maximum Number of Logs
Temperature
Material
Range
Unit of Measure
Accuracy
Resolution
31.5cm
18.3mm
0.5 kg
Titanium
Black Delrin®
Modbus RS-485, SDI-12,
4-20 mA
Modbus RS485 1 per sec
SDI-12 1 per sec
4-20 mA 1 per sec
5psi–500psi
-20°C–65°C
-40°C–65°C
3.6V lithium
8-36 VDC
Guaranteed - 5 yrs or 200,000 readings*
4MB, 380,000 data points**
Linear 1 per minute
Linear Average*** 1 per sec
Event*** 1 per sec
Linear 1 per minute
Linear Average*** 1 per 2 sec
Event*** 1 per sec
50
Titanium
-20°C to 65°C
Celsius or Fahrenheit
+/- 0.1ºC
0.01ºC
Conductivity
Sensor Type
Sensor Material
Range
Accuracy
Resolution
Parameters Supported*
Actual Conductivity
Specic Conductivity**
Salinity***
TDS
Resistivity
Density (Water)
Pressure/Level****
Sensor Type
Material
Range, Non-Vented
Range, Vented
Unit of Measure
Accuracy*¤*
Resolution
Max. Pressure
Burst Pressure
Balanced 4-electrode cell
PVC, Titanium
5µS/cm–100,000µS/cm
+/- 0.5% of reading plus 1 µS/cm from 5
µS/cm to 80,000 µS/cm
+/- 1.0% of reading from 80,000 µS/cm to
100,000 µS/cm
0.1µS/cm
µS,mS5µS/cm–100,000µS/cm
5µS/cm–100,000µS/cm
0–42PSU
ppt0–82ppt
10Ohms-cm–200,000Ohms-cm
0.98–1.14gm/cm3
Silicon Strain Gauge
Titanium
30, 100, 300, 500 psia
10.9, 60.1, 200.7, 341.3 m
5, 15, 30, 100, 300, 500 psig
33.5, 11, 21, 70, 210, 351 m
PSI, KPa, mm, cm, m, in., ft., Bar, mBar,
mmHg
@ 15°C +/- 0.05% Full Scale (FS)
0°C to 50°C +/- 0.1% FS
-20ºC to -1ºC and 51°C to 80ºC +/- 0.25% FS
+/- 0.005% FS or better
2X Range
3X range
* 1 reading = time and date plus all available parameters polled or logged from device.
** 1 data point = time and date plus one parameter logged.
*** External power supply or battery pack required when using event testing, recommended
when using linear average testing.
*Parametersderivedat25°CandActualConductivityrangeof5µS/cm–100,000µS/cmwitha
+/-0.05% plus 1µS/cm accuracy;
** Derived from Standard Methods 2510B.
*** Dened by the Practical Salinity Scale 1978 Standard Methods 2520B.
**** Real time Level compensation for Density of water.
*¤* Accuracy with 4-20 mA output option: +/- 0.25% FS.
Multi-Parameter