Instruction Manual αlpha-pH2000D
- 3 -
2 PRODUCT DESCRIPTION
2.1 Description of unit
The αlpha pH 2000D is used for measuring pH and temperature values. The pH values can be
measured using industrial differential & combination pH sensors. The temperature values can
be measured using 2-wire / 3-wire Pt100 / Pt1000 sensors or NTC300 thermistor. The
transmitter is available in a wall mounting enclosure. The controller / transmitter can be used
for applications such as water treatment and monitoring, galvanic-decontamination, chemical
processing, food processing, clean or wastewater control and neutralization processes.
This transmitter has many user-friendly and safety features which include:
• Push-button for calibration and sensor offset adjustment from the keypad.
• Built-in non-volatile memory to ensure that calibration and other information are not
erased if power supply fails.
• Menu-driven program that simplifies set-up.
• Automatic temperature compensation (ATC).
• Manual temperature compensation setting without the ATC probe, with independent
setting for calibration and process temperature.
• Two galvanically isolated current outputs 0/4...20mA.
• Two switching contacts as set-point relays.
• 0 to 2000 seconds time delay adjustment on all relays – minimizes false alarms.
• Separately adjustable high and low set-point hysteresis (dead bands) prevent chattering
of relays around the set points.
• Three control modes: limit controller, P controller and PI controller (P/PI controller as
pulse length or pulse frequency).
• Large dual display LCD for easy reading with clear multiple annunciators, alarm status,
operational and error messages.
• Separate alarm relay alerting you to set point limits exceeded for a certain time and if the
Pt100/Pt1000 or NTC300 thermistor wires are broken or disconnected during the ATC
function.
• Wash relay.
• Hold function to freeze output current (0/4...20mA) and release control relays.
• LED indicators signal control activities to visually monitor transmitter status from a
distance.
• Protection against electromagnetic interference.
• Back lit and UV light protected LC display.