Instruction Manual CON 510 Bench Conductivity/TDS Meter Tech nol ogyMadeEasy...68X090820 Rev. 0 11
Instruction Manual CON 510 7 4.6 Automatic Calibration (Only for Conductivity) In the automatic calibration mode, the CON 510 bench meter is capab
Instruction Manual CON 510 8 4.7 Manual Calibration (Conductivity/TDS) The meter accepts either single point or multi-point (up to 5 points) man
Instruction Manual CON 510 9 4.8 Temperature Calibration The Conductivity electrode (EC-CONSEN91W/ 35608-50) has a built-in temperature sensor f
Instruction Manual CON 510 10 5 MEASUREMENT The CON 510 bench meter measures up to 5 different measurement ranges with auto-ranging capability whi
Instruction Manual CON 510 11 5.2 Manual Temperature Compensation NOTE: For manual temperature compensation, you must deactivate the ATC mode to
Instruction Manual CON 510 12 5.4 Using Manual Ranging Function Although the CON 510 bench meter has an automatic ranging capability, you may al
Instruction Manual CON 510 13 6 MEMORY AND DATA INPUT FUNCTIONS 6.1 Memory Input The meter’s non-volatile memory can store and recall up to 50 da
Instruction Manual CON 510 14 7 SETUP FUNCTIONS The setup mode allows you to customize the meter’s setting to your individual preferences. The CON
Instruction Manual CON 510 15 SETUP . P5.0: Mode of Calibration Selection of Automatic or Manual Calibration (only in Conductivity mode) Sele
Instruction Manual CON 510 16 7.3 P2.0: Viewing Electrode Diagnosis Main program 2.0 shows the effective cell constant for each range being calib
Preface This manual serves to explain the use of the CON 510 bench meter. It functions in two ways, firstly as a step by step guide to help you to o
Instruction Manual CON 510 17 P3.2 Selecting °C or °F You can select between °C and °F as unit of measure for temperature readings. Meter default
Instruction Manual CON 510 18 P3.4 Setting the TDS factor (In TDS Mode) The concentration of salts dissolved in solution increases the conducti
Instruction Manual CON 510 19 P4.2 Adjusting the Normalization Temperature Your meter will normalize its conductivity measurements to a standard
Instruction Manual CON 510 20 7.6 P5.0 Mode of calibration The CON 510 bench meter allows you to perform automatic (only for Conductivity mode) o
Instruction Manual CON 510 21 7.7 P6.0 Selecting the cell constant This program allows you to select the appropriate cell constant of K = 0.1, 1.
Instruction Manual CON 510 22 7.8 P7.0: Resetting to factory default settings Program 7.0 allows you to reset all parameters to factory default s
Instruction Manual CON 510 23 8 PROBE CARE AND MAINTENANCE Keep the Conductivity/TDS probe clean. Rinse the probe twice, and gently swirl it whil
Instruction Manual CON 510 24 9 TROUBLE SHOOTING GUIDE Problem Possible Cause Solution No display A. AC outlet not switched on. A. Switch on
Instruction Manual CON 510 25 11 SPECIFICATIONS SPECIFICATIONS CON510 Conductivity Range 0 to 20.00, 200.0, 2000 µS/cm; 0 to 20.00, 200.0 mS/cm T
Instruction Manual CON 510 26 12 ACCESSORIES Replacement Meter and Accessories Item Description Eutech Instruments Order Code No. Oakton Inst
TABLE OF CONTENTS 1 INTRODUCTION 1 2 DISPLAY AND KEYPAD FUNCTIONS 1 2.1 Display 1 2.2 Keypad 2 3 PREPARATION 3 3.1 Conductivity Electrode
Instruction Manual CON 510 27 13 NOTE: CONDUCTIVITY AND TDS SOLUTIONS HAVE ±1% ACCURACY AT 25°C ADDENDUM 1: CALIBRATION TIPS You only need one
Instruction Manual CON 510 28 15 ADDENDUM 3: CALCULATING TEMPERATURE COEFFICIENTS To determine the temperature coefficient of your sample soluti
Instruction Manual CON 510 29 16 ADDENDUM 4: METER FACTORY DEFAULT SETTINGS Type Parameter Default Remarks P1.1 Viewing calibration solution
Instruction Manual CON 510 30 17 WARRANTY The CON 510 bench meter is supplied with a 3-year warranty from manufacturing defects and electrodes for
31 For more information on Eutech Instruments/ Oakton Instruments’ products, contact your nearest distribut
Instruction Manual CON 510 1 1 INTRODUCTION Thank you for selecting the CON510 bench meter. This meter is a microprocessor-based instrument that i
Instruction Manual CON 510 2 2.2 Keypad The splash-proof membrane tactile keypad allows easy key entry. Each button, when pressed, has a correspon
Instruction Manual CON 510 3 3 PREPARATION 3.1 Conductivity Electrode Information The CON510 bench meter is supplied with a Conductivity/TDS elec
Instruction Manual CON 510 4 3.4 Connecting the Electrode Holder The integral electrode holder serves as a handy holder for a
Instruction Manual CON 510 5 4 CALIBRATION 4.1 Important Information on Meter Calibration The CON 510 bench meter allows you to perform automat
Instruction Manual CON 510 6 Normalization Temperature: The factory default value for normalization temperature is 25 °C. If you need to normalize