Instruction Manual CyberScan pH 1100/ 2100
42
10.3.6 mV calibration data reset setup program P2.4
Use this program to erase all mV calibration data.
1. Enter mV setup mode as described in steps 1-4 in Section
10.3.1 Entering mV setup mode (page 37).
2. Use /MI key to scroll to program mV Cal Reset: P2.4.
See Figure 125.
3. Press ENTER to enter program mV Cal Reset: P2.4.
4. Press >/CAL key to toggle between mV Reset: Yes and No.
The selection will flash. See Figure 126.
5. Press ENTER to confirm selection.
If you selected Yes, the screen flashes mV being Reset,
Please wait... See Figure 127. The meter then returns to
setup mode.
If you select No, the screen flashes Reset Aborted and
then returns to the setup mode. See Figure 128. This is
the last mV setup program. Press /MR to go to the
previous program. Press /MI to go back to the first mV
setup program. Press EXIT to return to mV measurement
mode.
Reset Aborted
mV being Reset
Please wait. . .
mV Reset: Yes No
Select & Press ENTER
mV Cal Reset :P2.4
Press ENT. to Select
Figure 125
Figure 126
Figure 127
Figure 128