CyberScan pH1100/ 2100
59
15 TROUBLESHOOTING & ERROR MESSAGES
TROUBLESHOOTING
Problem Possible Cause Solution
Power ON but no display AC adapter not connected properly Connect AC adapter properly
Unstable Reading Electrode submersion in sample is
insufficient
Place electrode deeper into sample
Broken electrode Replace electrode
External noise causing instability in
reading.
Remove or switch off interfering motor
Dirty electrode Clean Electrode. Reactivate if necessary
Slow response Dirty electrode Clean Electrode. Reactivate if necessary
ERROR MESSAGES
LCD Display Indicates Cause Solution
Error:
EEPROM Read Error
Memory write error Unit is too old (> 10 years) Turn meter off and return to authorized
distributor if necessary
Error:
EEPROM write Error
Memory checksum
error
Hardware failure Turn meter off and return to authorized
distributor if necessary
Error:
DTR is set too high
DTR line is
disconnected
Hardware fault Turn meter off and return to authorized
distributor if necessary
Error:
Communication error
Frame/parity errors Turn meter off and return to authorized
distributor if necessary
Error:
Buffer Overflow
Received buffer
overflowing
Turn meter off and return to authorized
distributor if necessary
Error:
pH Offset Error
pH offset too much pH offset should not exceed
±1 pH from the display
Check electrode condition
Make sure buffer is fresh and correct
Recalibrate meter
Error:
pH Slope Error
Incorrect pH slope pH slope is out of allowable
limit
Check electrode condition
Make sure buffer is fresh and correct
Recalibrate meter
Error:
Ion Slope Error
(pH 2100 meter only)
Incorrect ion slope Average ion slope is not
within 10mV-75mV/decade
Check electrode condition
Make sure buffer is fresh and correct
Recalibrate meter
Error:
mV Offset too large
Offset too much Value input is not within
±150.0mV from the display
Recalibrate meter
Error:
Out of Range
Out of Range Value input exceeds the
specified range
Re-enter value
Error:
ATC offset too much
Offset of temp. probe
too much
Temp. calibration offset is
not within ±5
o
C (or ±9
o
F)
Re-enter value
Error:
Too much offset@7pH
Offset too much pH offset should not exceed
±1 pH from the display
(custom pH buffer)
Re-enter value
Error:
Slope out:75%-105%
Incorrect pH slope pH slope is not within 75% to
105% (custom pH buffer)
Re-enter value
Error: Pts too close
Or slope diff. sign
Incorrect slope
alignment
Slope is not negative or cal.
Points are not ±1 pH apart
(custom pH buffer)
Re-enter value