Instruction Manual CyberScan PC 5500/ 5000 Bench pH, Ion & Conductivity Meter/ Bench pH & Conductivity Meter (Also Applicable for CyberScan

Strany 2

Instruction Manual CyberScan PC5500/ 5000 5 2. Connect the electrode arm to the base. 3. Connect the power cable to the c

Strany 3 - PREFACE

Instruction Manual CyberScan PC5500/ 5000 95There are two ways to access the Conductivity Setup screen. From the Setup screen 1. T

Strany 4 - TABLE OF CONTENTS

Instruction Manual CyberScan PC5500/ 5000 96 Select from the options to the rightpHmVIonCondsetupE U T E C H INSTRUMENTSeditSTCondu

Strany 5

Instruction Manual CyberScan PC5500/ 5000 97To Set Sample ID# When this option is active, each time you touch print on the Measure

Strany 6 - 1 INTRODUCTION

Instruction Manual CyberScan PC5500/ 5000 98 Current UNITConductivitySalinityResistivityTotal Dissolved SolidsSelect Conductivity

Strany 7 - 2 UNPACKING THE METER

Instruction Manual CyberScan PC5500/ 5000 999.2 Select Cell Constant Current CONSTANTSelect Cond Cell ConstantSTenter Use t

Strany 8 - 3 GETTING STARTED

Instruction Manual CyberScan PC5500/ 5000 100This screen allows you to select the appropriate cell constant of the conductivity cell

Strany 9 - Back Panel of PC 5500/ 5000

Instruction Manual CyberScan PC5500/ 5000 1019.3 Select Reference Temperature Current TEMPERATURESelect Cond Reference TempSTen

Strany 10

Instruction Manual CyberScan PC5500/ 5000 102To Select Reference Temperature This screen allows you to select the reference tempera

Strany 11

Instruction Manual CyberScan PC5500/ 5000 1039.4 Set Temperature Coefficient Set Cond Temperature CoeffCurrent COEFF Touch

Strany 12

Instruction Manual CyberScan PC5500/ 5000 104The temperature coefficient is a value that reflects the degree to which the sample’s c

Strany 13

Instruction Manual CyberScan PC5500/ 5000 63.2 pH Electrodes & Ion Specific Electrodes (PCD5500 & PC5500 only This meter a

Strany 14

Instruction Manual CyberScan PC5500/ 5000 105You are about to replatinize an electrodePlace the electrode in replarinizing solution

Strany 15 - 11:11 am

Instruction Manual CyberScan PC5500/ 5000 106Replatinization is the process of replacing the platinum on the surfaces of the 2-cell

Strany 16

Instruction Manual CyberScan PC5500/ 5000 107 To Set Alarm Limits This option allows you to set alarm limits for the Conductivity m

Strany 17

Instruction Manual CyberScan PC5500/ 5000 108To Set Data Storage Criteria This screen allows you to select what criteria are stored

Strany 18

Instruction Manual CyberScan PC5500/ 5000 1099.6 Set Significant Digits Set Cond Significant DigitsCurrent SIG DIGITS Select desi

Strany 19

Instruction Manual CyberScan PC5500/ 5000 110This option allows you to select the number of significant figures that will be used wh

Strany 20

Instruction Manual CyberScan PC5500/ 5000 111 10 pH OPERATION MeasureBuffers  Touch to measure sample or Touch to stand

Strany 21 - System Setup

Instruction Manual CyberScan PC5500/ 5000 112In this mode, you will able to measure the pH of a wide variety of samples. Before meas

Strany 22

Instruction Manual CyberScan PC5500/ 5000 113MeasureBuffers  Touch to measure sample or Touch to standardizemeas stdmeasst

Strany 23

Instruction Manual CyberScan PC5500/ 5000 114To Standardize the meter with Auto Buffer Recognition 1. Touch STD on the pH measure s

Strany 24 - Language

Instruction Manual CyberScan PC5500/ 5000 7 3. Rinse and blot-dry (don’t wipe) electrodes between each measurement. Rinse electro

Strany 25

Instruction Manual CyberScan PC5500/ 5000 115 Buffers  To u c h to delete previous buffers Insert electrode into buffer an

Strany 26 - Current MM/DD/YY

Instruction Manual CyberScan PC5500/ 5000 116To standardize the meter with Manual Buffer Recognition 1. Touch std on the pH measure

Strany 27

Instruction Manual CyberScan PC5500/ 5000 11710.2 pH Measurements MeasureBuffers  Touch to measure sample or Touch to

Strany 28

Instruction Manual CyberScan PC5500/ 5000 118The measure screen provides readout of the current sample measurement. You can use this

Strany 29 - Current 24HR

Instruction Manual CyberScan PC5500/ 5000 11911 mV OPERATION This mode is used to measure oxidation/ reduction potential (ORP/redox

Strany 30 - Beeper Status

Instruction Manual CyberScan PC5500/ 5000 12011.1 mV Measurement In the mV mode, you will be able to make measurements in either a

Strany 31

Instruction Manual CyberScan PC5500/ 5000 121 12 ION OPERATION Measure - Ionstdppmchannel 117STNDSLast std: Jan 15 @ 11:11 am10.0

Strany 32 - Procedural Level

Instruction Manual CyberScan PC5500/ 5000 122This CyberScan meter offers you unparalled versatility with respect to ion analysis. Yo

Strany 33

Instruction Manual CyberScan PC5500/ 5000 123clearStandardize - IonsetupTouch to measure sampleormeasTouch to access standardize

Strany 34 - Print Configuration

Instruction Manual CyberScan PC5500/ 5000 124There are two direct reading methods. Direct Reading with standards and Direct Reading

Strany 35

Instruction Manual CyberScan PC5500/ 5000 83.3 Conductivity Probes This meter allows you to use two types of conductivity cells: th

Strany 36 - Baud Rate

Instruction Manual CyberScan PC5500/ 5000 125Measuring ion concentration in the Direct Reading with Standards Mode With Auto Read o

Strany 37 - Number of Bits

Instruction Manual CyberScan PC5500/ 5000 126Standardize - IonSTND Insert electrode into standard and stir Touch to standardiz

Strany 38 - Stop Bits

Instruction Manual CyberScan PC5500/ 5000 127Direct Reading with Blank Offset This method presents you with the ability to measure

Strany 39 - Current PARITY

Instruction Manual CyberScan PC5500/ 5000 128If you touched STND 8. Use the keypad to enter the value of the standard. 9. Touch e

Strany 40 - Operator

Instruction Manual CyberScan PC5500/ 5000 129clearStandardize - IonsetupTouch to measure sampleormeasTouch to access standardize

Strany 41

Instruction Manual CyberScan PC5500/ 5000 130The incremental methods which include known addition, known subtraction, analate additi

Strany 42 - Display Contrast

Instruction Manual CyberScan PC5500/ 5000 131Measure - IonSTND Touch to access standardize mode stdKAstdLast std: Jan 15 @ 1

Strany 43

Instruction Manual CyberScan PC5500/ 5000 132clearStandardize - IonsetupTouch to measure sampleormeasTouch to access standardiz

Strany 44

Instruction Manual CyberScan PC5500/ 5000 133Measuring Ion Concentration in Samples Using the Incremental Methods When performing k

Strany 45 - Reset to Factory Defaults

Instruction Manual CyberScan PC5500/ 5000 134To take a measurement 1. Immerse the ion selective electrode in a known volume of the

Strany 46 - 6 pH SETUP

Instruction Manual CyberScan PC5500/ 5000 93. Rinse and blot-dry (don’t wipe) probe between each measurement. Rinse probe with dist

Strany 47

Instruction Manual CyberScan PC5500/ 5000 13512.4 Known Subtraction Method The procedure for known subtraction is very similar to

Strany 48

Instruction Manual CyberScan PC5500/ 5000 13610. Use the keypad to enter the concentration of the standard solution added to the sam

Strany 49 - Manual ID#

Instruction Manual CyberScan PC5500/ 5000 13712.5 Analate addition The analate addition method is useful when your sample is at an

Strany 50

Instruction Manual CyberScan PC5500/ 5000 13810. Use the keypad to enter the concentration of the standard solution added to the sam

Strany 51 - Sequential ID#

Instruction Manual CyberScan PC5500/ 5000 13912.6 Analate Subtraction The analate subtraction method is useful when no ion specifi

Strany 52

Instruction Manual CyberScan PC5500/ 5000 14010. Use the keypad to enter the concentration of the standard solution added to the sam

Strany 53 - and then touch to accept

Instruction Manual CyberScan PC5500/ 5000 14113 CONDUCTIVITY OPERATION Measure - ConductivitystdµS/cmchannel 10.010STNDLast std:

Strany 54

Instruction Manual CyberScan PC5500/ 5000 142This mode allows you to measure the conductivity of solutions. Conductance is a value a

Strany 55 - 1.78 3.65

Instruction Manual CyberScan PC5500/ 5000 14313.1 Conductivity Standardization   Touch to measure sample orTouch to access

Strany 56

Instruction Manual CyberScan PC5500/ 5000 144To “standardize” your conductivity probe in the Advanced Level of Operation 1. Access

Strany 57 - Current RECOGNITION

Instruction Manual CyberScan PC5500/ 5000 104 USING THE METER Touch anywhere to resumeE U T E C H INSTRUMENTS11:11 am

Strany 58 - Current MODE

Instruction Manual CyberScan PC5500/ 5000 145To “standardize” your Conductivity Probe in the Basic Level of Operation At the Basic

Strany 59 - Current CRITERIA

Instruction Manual CyberScan PC5500/ 5000 14613.2 Measurement Measure - ConductivitystdµS/cmchannel 10.010STNDLast std: Jan 15 @

Strany 60

Instruction Manual CyberScan PC5500/ 5000 147Once an exact cell constant has been established, the meter is ready for sample measure

Strany 61 - Current DEFAULT

Instruction Manual CyberScan PC5500/ 5000 14814 CLEANING The touch screen should be kept as clean as possible to preserve optical

Strany 62

Instruction Manual CyberScan PC5500/ 5000 14915 TROUBLE SHOOTING Your CyberScan meter contains many error messages to provide aid s

Strany 63 - Current ISO POINT

Instruction Manual CyberScan PC5500/ 5000 15016 COMPLIANCE Notice of compliance Warning: this meter generates uses and can radiate

Strany 64

Instruction Manual CyberScan PC5500/ 5000 151Figure 1Figure 217 APPENDIX 17.1 Determining Isopotential Points Experimentally The i

Strany 65 - Current LIMITS

Instruction Manual CyberScan PC5500/ 5000 152 17.2 Data Management Your CyberScan meter is equipped with an RS-232 output port for

Strany 66

Instruction Manual CyberScan PC5500/ 5000 15317.3 Factory Default Settings The following is the list of factory default settings f

Strany 67

Instruction Manual CyberScan PC5500/ 5000 154 Mode Screen Default Setting ISE Display configuration Print configuration Data co

Strany 68

Instruction Manual CyberScan PC5500/ 5000 114.1 Touch screen Operation This new CyberScan research benchtop meter operates with a

Strany 69 - Current INTERVAL

Instruction Manual CyberScan PC5500/ 5000 155 17.4 pH Theory Since its introduction by the Danish Chemist Sorensen in 1909, pH mea

Strany 70

Instruction Manual CyberScan PC5500/ 5000 156At 25 °C in pure water, the concentration of hydronium ions is extremely small, 1 X 10-

Strany 71

Instruction Manual CyberScan PC5500/ 5000 157pH Measurement The pH value of a sample can be determined in several ways. These inclu

Strany 72

Instruction Manual CyberScan PC5500/ 5000 158 Temperature 4.01 6.86 9.18 0 10 20 25 30 40 50 4.003 3.998 4.002 4.008 4.015 4.03

Strany 73 - touch to accept

Instruction Manual CyberScan PC5500/ 5000 159Reference electrodes typically consist of three elements: an internal reference electro

Strany 74 - Example of X.XXX resolution

Instruction Manual CyberScan PC5500/ 5000 16017.5 ISE Theory Ion selective electrodes (ISE) respond to ionic activity or concentra

Strany 75

Instruction Manual CyberScan PC5500/ 5000 161This form of the Nernst equation states that the electrode potential varies directly wi

Strany 76

Instruction Manual CyberScan PC5500/ 5000 162Known addition and Known Subtraction With the Known Addition and Known Subtraction met

Strany 77 - Stored Data

Instruction Manual CyberScan PC5500/ 5000 163Analate Addition and Analate Subtraction In the Analate methods, the initial millivolt

Strany 78

Instruction Manual CyberScan PC5500/ 5000 16417.6 Conductivity Theory Conductance is a quantity associated with the ability o prima

Strany 79

Instruction Manual CyberScan PC5500/ 5000 124.2 Choosing a channel (for PCD5500 & PC5500 only) Select from the options t

Strany 80

Instruction Manual CyberScan PC5500/ 5000 165Thus comes the term specific conductance or conductivity. The relationship between con

Strany 81 - 7 mV SETUP

Instruction Manual CyberScan PC5500/ 5000 166Conductivity Measurement Conductivity electrodes consist of glass or epoxy bodies in w

Strany 82

Instruction Manual CyberScan PC5500/ 5000 167 The four-cell electrode Traditionally, conductivity measurements were made with a “tw

Strany 83

Instruction Manual CyberScan PC5500/ 5000 168Since temperature effects the conductivity measurement so profoundly, the usual practic

Strany 84

Instruction Manual CyberScan PC5500/ 5000 169 18 METER SPECIFICATIONS Model Details PCD 5500 PC 5500 PC 5000 Display 640X480 dig

Strany 85

Instruction Manual CyberScan PC5500/ 5000 170 Model Details PCD 5500 PC 5500 PC 5000 Conductivity Mode Cell constants range

Strany 86

Instruction Manual CyberScan PC5500/ 5000 17119 ACCESSORIES Consult your Authorized Distributors for these items and other range of

Strany 87 - Example of X resolution

Instruction Manual CyberScan PC5500/ 5000 17219.2 General Purpose pH / ORP Electrodes SP CODE NO. DESCRIPTION TYPICAL APPLICATION

Strany 88 - Ion Setup (channel 1)

Instruction Manual CyberScan PC5500/ 5000 17319.3 Specialty pH Electrode NP CODE NO. DESCRIPTION TYPICAL APPLICATIONS EC-620-297

Strany 89

Instruction Manual CyberScan PC5500/ 5000 17419.4 pH Buffer Solutions & Single Use Sachets SP CODE NO. DESCRIPTION EC-BU-1BT

Strany 90

Instruction Manual CyberScan PC5500/ 5000 13The CyberScan PCD5500/ PC5500 are multi-channel meters. With this meter you can switch f

Strany 91 - Ion Method

Instruction Manual CyberScan PC5500/ 5000 17520 WARRANTY Eutech Instruments supplies the bench meters with a 3-year warranty and

Strany 93 - Measurement Units

For more information on Eutech Instruments products, contact your nearest Eutech Instruments distributor or visit

Strany 94

Instruction Manual CyberScan PC5500/ 5000 14savemodeenterclosemoreback4.3 Button Functions The touch screen of your CyberScan bench

Strany 96

Instruction Manual CyberScan PC5500/ 5000 15prevnextprintmeassetupeditclearstdBSBSdeleteSThis button accesses the standardization sc

Strany 97 - 8.5 Set Significant Digits

Instruction Manual CyberScan PC5500/ 5000 165 SYSTEM SETUP SYSTEM SETUP OPTIONS- Select Language- Set - Set - Set - Set - Set - S

Strany 98

Instruction Manual CyberScan PC5500/ 5000 175.1 Access system setup SYSTEM SETUP OPTIONS- Select Language- Set - Set - Set - S

Strany 99 - 9 CONDUCTIVITY SETUP

Instruction Manual CyberScan PC5500/ 5000 18The system setup function allows you to customise the meter display options to meet your

Strany 100 - INSTRUMENTS

Instruction Manual CyberScan PC5500/ 5000 195.2 Select Language Current LANGUAGEEnglishFrenchGermanSpanishSet LanguageSTenterU

Strany 101

Instruction Manual CyberScan PC5500/ 5000 20This option allows you to choose the language in which all prompts and directions will a

Strany 102

Instruction Manual CyberScan PC5500/ 5000 215.3 Set Date Set DateCurrent MM/DD/YYclearTouch clear to delete current dateUse nu

Strany 103 - Conductivity Units

Instruction Manual CyberScan PC5500/ 5000 22This screen can be used to set the present date which will be displayed on the measure s

Strany 104 - Cond Cell Constant

Instruction Manual CyberScan PC5500/ 5000 235.4 Set Time Set TimeCurrent 12HR:clearTouch to delete current timeUse to inp

Strany 105

Instruction Manual CyberScan PC5500/ 5000 24This screen can be used to set the present time which will be displayed on the measure s

Strany 106 - Cond Reference Temp

PREFACE Thank you for selecting a Eutech Instruments CyberScan Series 5000 bench meter. This meter measures pH, conductivity and ion (PCD5500 &

Strany 107

Instruction Manual CyberScan PC5500/ 5000 255.5 Set Beeper Status Current BEEPER STATUS- After STABLE Readin - A

Strany 108 - 0.00%/°C

Instruction Manual CyberScan PC5500/ 5000 26This screen allows you to turn on or off the beeper. You may choose to have an audible s

Strany 109

Instruction Manual CyberScan PC5500/ 5000 275.6 Select procedural level enter12Current LEVELSelect Procedural LevelLEVEL 1LEV

Strany 110 - Replatinize

Instruction Manual CyberScan PC5500/ 5000 28 This selection screen allows you to choose the amount of information that you want to h

Strany 111

Instruction Manual CyberScan PC5500/ 5000 295.7 Set Print Configuration Current PRINT CONFIGURATION- Baud Rate

Strany 112

Instruction Manual CyberScan PC5500/ 5000 30You can adjust the print configuration of this meter from this screen. The configuration

Strany 113

Instruction Manual CyberScan PC5500/ 5000 315.8 Set Baud Rate This configuration option will control the speed at which the data w

Strany 114 - touch to accept

Instruction Manual CyberScan PC5500/ 5000 32 5.9 Set Number of Bits To Set Number of Bits 1. Access the Set number of Bits scre

Strany 115

Instruction Manual CyberScan PC5500/ 5000 33 5.10 Set Stop Bits To Set Stop Bits 1. Access the Set Stop Bits screen from the Se

Strany 116 - 4 7 10

Instruction Manual CyberScan PC5500/ 5000 34 5.11 Set Parity To Set Parity 1. Access the Set Parity screen from the Set Print C

Strany 117

TABLE OF CONTENTS 1 INTRODUCTION 1 1.1 Introducing the CyberScan Series 1 2 UNPACKING THE METER 2 3 GETTING STARTED 3 3.1 Connectors (PCD 5

Strany 118 - Standardize

Instruction Manual CyberScan PC5500/ 5000 35 5.12 Set Operator Set OperatorCurrent OperatorA B C

Strany 119

Instruction Manual CyberScan PC5500/ 5000 36This option allows you to identify the user of the meter. This information can be saved

Strany 120

Instruction Manual CyberScan PC5500/ 5000 375.13 Set Display Contrast Set Display ContrastTouch or to adjust contrast a

Strany 121

Instruction Manual CyberScan PC5500/ 5000 38This option allows you to change the contrast on the screen to improve the readability o

Strany 122

Instruction Manual CyberScan PC5500/ 5000 395.14 Display Meter Information This screen displays the model number, serial number an

Strany 123

Instruction Manual CyberScan PC5500/ 5000 40 5.15 Reset to Factory Defaults This screen allows you to reset all functions and setup

Strany 124 - 11 mV OPERATION

Instruction Manual CyberScan PC5500/ 5000 41 6 pH SETUP pH SETUP OPTIONS- Set - Select - Select - Se

Strany 125

Instruction Manual CyberScan PC5500/ 5000 42 6.1 Access pH setup There are two ways to access the pH Setup screen. From the se

Strany 126 - Measure - Ion

Instruction Manual CyberScan PC5500/ 5000 43 From the pH measure screen 1. Touch pH on the main screen to access the pH mode. Touc

Strany 127

Instruction Manual CyberScan PC5500/ 5000 44 6.2 Set Sample ID# Manual ID#enterclearA B C

Strany 128 - Standardize - Ion

9 CONDUCTIVITY SETUP 94 9.1 Access the Conductivity Setup 95 9.2 Select Cell Constant 99 9.3 Select Reference Temperature 101 9.4 Set Tempe

Strany 129

Instruction Manual CyberScan PC5500/ 5000 45When this option is active, each time you touch print on the Measure screen the pH valu

Strany 130

Instruction Manual CyberScan PC5500/ 5000 46 Set pH Sample ID#Sequential ID#clearTouch to delete current IDUse to input

Strany 131

Instruction Manual CyberScan PC5500/ 5000 47Sequential ID# Assignment 1. Access the Set Sample ID# screen from the pH (mV, Ion, Co

Strany 132

Instruction Manual CyberScan PC5500/ 5000 486.3 Select Buffer Group View pH Buffer GroupCurrent Buffer GroupcustomNISTUSAUse

Strany 133

Instruction Manual CyberScan PC5500/ 5000 49This setup option allows you to select from 3 different buffer groups each containing 5

Strany 134 - 12.2 Incremental Methods

Instruction Manual CyberScan PC5500/ 5000 50View pH Custom Buffer GroupCurrent BUFFER GROUPCUSTOMUse to display desired buffer gr

Strany 135

Instruction Manual CyberScan PC5500/ 5000 51To Set pH Custom Buffer Group This option allows you to create a custom buffer group of

Strany 136 - 12.3 Known Addition Method

Instruction Manual CyberScan PC5500/ 5000 526.5 Select Buffer Recognition This option allows you to select automatic buffer recogn

Strany 137 - Known Addition Result

Instruction Manual CyberScan PC5500/ 5000 53 6.6 Select Auto Read Mode You can use this meter when the Auto Read Mode function is

Strany 138

Instruction Manual CyberScan PC5500/ 5000 54 6.7 Set pH Stability Criteria Current CRITERIAFastSlowMediumSetpH Stability Criteria

Strany 139

Instruction Manual CyberScan PC5500/ 5000 1 1 INTRODUCTION 1.1 Introducing the CyberScan Series Thank you for selecting a Eutech

Strany 140

Instruction Manual CyberScan PC5500/ 5000 55This setup screen allows you to determine how quickly the meter will respond to electrod

Strany 141

Instruction Manual CyberScan PC5500/ 5000 56 6.8 Set Default Temperature Set pH Default TemperatureCurrent DEFAULT Touch to d

Strany 142

Instruction Manual CyberScan PC5500/ 5000 57It is a well known fact that pH is a temperature dependent measurement. The factory defa

Strany 143

Instruction Manual CyberScan PC5500/ 5000 586.9 Set Isopotential Point Set pH Isopotential PointCurrent ISO POINT Touch to

Strany 144

Instruction Manual CyberScan PC5500/ 5000 59The Isopotential point is the millivolts reading for an electrode at which temperature h

Strany 145

Instruction Manual CyberScan PC5500/ 5000 606.10 Set Alarm Limits Current LIMITSpH AlarmpH MinimumpH MaximumSet pH LimitsSTeditOF

Strany 146 - 13 CONDUCTIVITY OPERATION

Instruction Manual CyberScan PC5500/ 5000 61This option allows you to set alarm limits for the pH measuring mode. If the pH value of

Strany 147

Instruction Manual CyberScan PC5500/ 5000 62Current PRINT CRITERIA- Date/Time/Channel - Sample ID#

Strany 148 - Measure - Conductivity

Instruction Manual CyberScan PC5500/ 5000 63This screen allows you to select which criteria are printed with the measurement when yo

Strany 149

Instruction Manual CyberScan PC5500/ 5000 646.12 Set Print Interval Set pH Print IntervalCurrent INTERVALtimedenter1

Strany 150

Instruction Manual CyberScan PC5500/ 5000 22 UNPACKING THE METER The following is a listing of what you should have received with

Strany 151

Instruction Manual CyberScan PC5500/ 5000 65For stable reading printed In this mode, data is printed every time the meter recognise

Strany 152

Instruction Manual CyberScan PC5500/ 5000 66Current DATA STORAGE CRITERIA- Date/Time/ Channel - Sample ID#

Strany 153 - 14 CLEANING

Instruction Manual CyberScan PC5500/ 5000 67 This screen allows you to select what criteria are stored in the meter’s memory with th

Strany 154 - 15 TROUBLE SHOOTING

Instruction Manual CyberScan PC5500/ 5000 686.14 Set Display Resolution Set pH Display ResolutionCurrent RESOLUTIONSelect desi

Strany 155 - 16 COMPLIANCE

Instruction Manual CyberScan PC5500/ 5000 69This mode allows you to set the display resolution that you desire on the screen. You ha

Strany 156 - 17 APPENDIX

Instruction Manual CyberScan PC5500/ 5000 70Current DISPLAY CONFIGURATION- Last Standardization- Date- Time- measurement channel- sa

Strany 157

Instruction Manual CyberScan PC5500/ 5000 71January 15, 2002 11:11amID# 00000auto buffer ONATC 25.0°Csl

Strany 158

Instruction Manual CyberScan PC5500/ 5000 72 6.16 View Stored Data Data POINTSSample IDDateOperatorView Stored DataSTenter Use

Strany 159

Instruction Manual CyberScan PC5500/ 5000 73This meter has memory capacity of up to 250 data points (1000 data points for PCD5500).

Strany 160

Instruction Manual CyberScan PC5500/ 5000 74To sort by Date 1. Access the Date sort option from the View Stored Data screen. 2.

Strany 161

Instruction Manual CyberScan PC5500/ 5000 33 GETTING STARTED 3.1 Connectors (PCD 5500 only) 1. Review the layout and arrangement

Strany 162 - E* - (pH)

Instruction Manual CyberScan PC5500/ 5000 75If an operator ID is entered and no data points are stored with that operator ID, you wi

Strany 163

Instruction Manual CyberScan PC5500/ 5000 76 7 mV SETUP mV SETUP OPTIONS- Set - Set - Set - Set - Set - Set - View Sample ID#Al

Strany 164

Instruction Manual CyberScan PC5500/ 5000 77 7.1 Access mV Setup There are two ways to access the mV Setup screen. From the

Strany 165 - E* ± (log c)

Instruction Manual CyberScan PC5500/ 5000 78 From the Setup screen 1. Touch setup on the mV Measure screen. The mV Setup screen

Strany 166

Instruction Manual CyberScan PC5500/ 5000 79To Set Sample ID# When this option is active, each time you touch print on the Measure

Strany 167 - (V +V )10 - V

Instruction Manual CyberScan PC5500/ 5000 80To Set Print Interval You have three options for setting the print interval: manual pri

Strany 168 - ( )

Instruction Manual CyberScan PC5500/ 5000 817.2 Set Display Resolution Set mV Display ResolutionCurrent RESOLUTIONSelect desire

Strany 169

Instruction Manual CyberScan PC5500/ 5000 82 To Set Display Resolution This mode allows you to set the display resolution that you

Strany 170 - megohmcm

Instruction Manual CyberScan PC5500/ 5000 838 ION SETUP ISE SETUP OPTIONS- Set - Select - Select - Set -

Strany 171

Instruction Manual CyberScan PC5500/ 5000 84 Select from the options to the rightpHmVIonCondsetupE U T E C H INSTRUMENTS Select fr

Strany 172

Instruction Manual CyberScan PC5500/ 5000 4 Back Panel of PC 5500/ 5000

Strany 173

Instruction Manual CyberScan PC5500/ 5000 85From the Ion Measure screen 4. Touch Ion on the main screen to access the Ion mode. Tou

Strany 174 - 18 METER SPECIFICATIONS

Instruction Manual CyberScan PC5500/ 5000 86 8.2 Select Ion Method To Set Sample ID# When this option is active, each time you to

Strany 175

Instruction Manual CyberScan PC5500/ 5000 87Direct Reading with Standards The direct reading method is the most common method of io

Strany 176 - 19 ACCESSORIES

Instruction Manual CyberScan PC5500/ 5000 88Analate Subtraction The analate subtraction method is particularly useful in the event

Strany 177

Instruction Manual CyberScan PC5500/ 5000 89To Select Auto Read Mode You can use this meter when the Auto Read Mode function is act

Strany 178

Instruction Manual CyberScan PC5500/ 5000 908.4 Select Electrode Type This screen allows you to select the type of electrode that

Strany 179

Instruction Manual CyberScan PC5500/ 5000 91To Set Isopotential point The Isopotential point is the millivolts reading for an elect

Strany 180 - 20 WARRANTY

Instruction Manual CyberScan PC5500/ 5000 928.5 Set Significant Digits ppmchannel 1Example of 1 significant digitSet Ion Significa

Strany 181

Instruction Manual CyberScan PC5500/ 5000 93 This option allows you to select the number of significant figures that will be used wh

Strany 182

Instruction Manual CyberScan PC5500/ 5000 94Conductivity Setup (channel 1)CONDUCTIVITY SETUP OPTIONSeditST- Set- Select - Select - S