Instruction Manual TN-100/ T-100
14
For continuous measurement:
1. Make sure the meter is sitting on a flat and level surface and is in
measurement mode. The display shows the last measured value or
[STbY].
2. Place the sample vial in the sample well.
3. Press READ/ENTER key and hold. See Figure 13.
4. Wait for the reading to stabilize before rotating the sample vial.
NOTE:
During continuous measurement, the display is
updated every 2 seconds. The displayed reading
may not be the actual turbidity value. For accurate
measurement, use single-shot measurement.
5. You can rotate the sample vial for
indexing purpose. See Section 8.5 -
Indexing a Vial on page 20 for more
information.
6. Once you release the READ/ENTER
key, the meter will automatically
perform a single-shot measurement.
NOTES:
1. After a measurement is completed and the
display is updated, a 4-second recovery time
occurs before the meter can perform another
function. If any key is pressed during the recovery
time, the meter will perform the corresponding
action at the end of the recovery period.
2. When performing a measurement, if the meter
detects stray light exceeding the amount
equivalent to 0.02 NTU in the low range, the
measurement is immediately aborted and an
error message [ERR 8] is displayed. Ensure the
vial sits properly into the sample well. Press the
ENTER/READ key to re-do the measurement.
SAMPLE
Unknown
NTU
Place vial
into the
sample well
Press and hold
NTU
READ
ENTER
You ca n tur n
the vial for
indexing
Figure 13: Continuous
measurement