Instruction Manual TN-100/ T-100
15
4 TROUBLESHOOTING GUIDE
The TN-100/ T-100 turbidimeter routinely performs self-diagnostics, and
will automatically generate messages to provide you with specific
diagnostic information. These messages are for your use and do not
indicate a reduction in the performance of the instrument or a failure of any
component in the instrument, unless otherwise stated in this list.
LCD
Message
Description Correction Actions
ERR 1 Calibration Error. The meter is
unable to recognize the 800 NTU
Calibration Standard.
Make sure you use the correct
800 NTU calibration standard. *
ERR 2 Calibration Error. The meter is
unable to recognize the 100 NTU
Calibration Standard.
Make sure you use the correct
100 NTU calibration standard. *
ERR 3 Calibration Error. The meter is
unable to recognize the 20.0 NTU
Calibration Standard.
Make sure you use the correct
20.0 NTU calibration standard. *
ERR 4 Calibration Error. The meter is
unable to recognize the 0.02 NTU
Calibration Standard.
Make sure you use the correct
0.02 NTU calibration standard. *
ERR 5 Calibration Error. There is not
sufficient signal to achieve the
appropriate resolution in the 0-
1000NTU range.
Re-do calibration with all 4
standards provided. *
ERR 6 General Calibration Failure. There
is not sufficient signal to achieve
the appropriate resolution in the 0-
100NTU range.
Re-do calibration with all 4
standards provided. *
ERR 7 General Calibration Failure. There
is not sufficient signal to achieve
the appropriate resolution in the 0-
20 NTU range.
Re-do calibration with all 4
standards provided. *
ERR 8 Excessive stray light detected. Make sure the vial is fully
snapped in the sample well.
ERR 9 Lamp Failure. Return unit.
Or When Turbidity value is above the
measurement range(>1000NTU)
Dilute sample. See Section 9.7
on page 23.
Low battery indication. The
batteries need to be replaced.
Replace batteries.
* If an error message appears, take the
appropriate corrective action and re-do the
desired procedure. If the problem persists,
contact Eutech Instruments or your dealer.
See Sections on Warranty and Return of Items on page 24.