Transmitter Cond/Res 2-Channel M300 26
© 05/07 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter Cond/Res 2-Channel M300
Printed in Switzerland 52 121 318
7.3.1 One-Point Sensor Temperature Calibration
See 7.3 on how to enter the Temperature Calibration menu. Slope or Offset can be selected with
the 1 point calibration. Select Slope to recalculate the Slope factor M (Multiplier) or Offset to
recalculate the offset calibration factor A (Adder).
Enter the value for Point 1 and press [ENTER].
The newly calculated value – either M or A – is displayed. Select Yes to save the new calibration
values and the successful Calibration is confirmed on the display.
7.3.2 Two – Point Sensor Temperature Calibration
See 7.3 on how to enter the Temperature Calibration menu. Select 2 point as calibration Type.
Enter the value for Point 1 and press [ENTER].
Enter the value for Point 2 and press [ENTER].
The newly calculated values M and A are displayed. Select Yes and press [ENTER] to save the
new calibration values and the successful Calibration is confirmed on the display.