11.2 Calibrate
(PATH: Menu/Service/Calibrate)
This menu has the options to calibrate the transmitter and the analog outputs and also allows
the unlocking of calibration functionality.
11.2.1 Calibrate Meter
The M300 transmitter is factory calibrated within specifications. It is not normally necessary to
perform meter re-calibration unless extreme conditions cause an out of spec operation shown by
Calibration Verification. Periodic verification/re-calibration may also be necessary to meet Q.A.
requirements.
The meter is equipped with five (5) internal ranges of measurement on each channel. Each
resistance range and temperature is calibrated separately, with each resistance range consisting
of a two-point calibration and the temperature using a three-point calibration.
Below is a table showing the resistance values for all calibration ranges.
Range: Point 1 Point 2 Point 3
Resistivity 1 1.0 Mohms 10.0 Mohms –
Resistivity 2 100.0 Kohms 1.0 Mohms –
Resistivity 3 10.0 Kohms 100.0 Kohms –
Resistivity 4 1.0 Kohms 10.0 Kohms –
Resistivity 5 100 Ohms 1.0 Kohms –
Temperature 1000 Ohms 3.0 Kohms 66 Kohms
It is recommended that both calibration and verification be performed using the M300 Calibrator
Module Accessory (refer to accessory list, in section 15). Instructions on the use of this acces-
sory are provided with the calibrator module.