Transmitter Cond/Res 2-Channel M300 61
© 05/07 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter Cond/Res 2-Channel M300
Printed in Switzerland 52 121 318
17 Default tables
Alarm Relay 2
Diagnostics off
Power Failure off
Software Failure off
Delay 1 sec
Hysteresis 0
State inverted
Clean Relay 1
Hold Mode NA
Interval 0 Hrs
Clean Time 0 Sec
State normal
Delay 0
Hysteresis 0
Language English
Passwords Administrator 00000
Operator 00000
All Relays
(unless otherwise specified)
Delay 10 sec
Hysteresis 5 %
State normal
Hold mode NA
Lockout (on/off) no = off
Channel A Measurement a Resistivity Ω-cm
Measurement b temperature °C
Channel B (Dual Channel) Measurement c Resistivity Ω-cm
Measurement d Temperature °C
Cal constants Cond/Res M=0.1, A=0.0
Temperature M=1.0, A=0.0
Analog Out 1 a – Cond/Resistivity MΩ-cm
2 b – temperature °C
3 c – Cond/Resistivity MΩ-cm
4 d – Temperature °C
All analog out Mode 4 – 20 mA
Type normal
Alarm off
Hold mode last value
Conductivity/Resistivity Value 4 mA 10 MΩ-cm
Value 20 mA 20 MΩ-cm
Temperature Value 4 mA 0 °C
Value 20 mA 100 °C
Set point 1 Measurement a
Type off
Value 0
Relay 3 Set Point 1
Set point 2 Measurement b = single channel
c = dual channel
Type Off
Value 0
Relay 4 Set Point 2
Resolution Temperature 0.1 °C
Cond/Res 0.01 Ω-cm