P3.2: Selecting number of pH calibration points
Program P3.2 lets you select the number of calibration
points that appear in pH calibration mode: 2, 3, 4, or 5.
The meter will automatically exit calibration mode after
you have calibrated to your selected number of points.
From measurement mode.
1. Press the MODE key to select pH measurement
mode.
2. Press SETUP key to enter Set Up mode.
3. Press the or keys to scroll through su/jointfilesconvert/267593/bgroups
until you view parameter P3.0.
4. Press the ENTER key twice to select parameter 3.2.
5. Press the or keys to select 2, 3, 4, or 5 point
pH calibration.
6. Press the ENTER key to confirm selection and to
return to the su/jointfilesconvert/267593/bgroup menu. Press CAL/MEAS
key to return to measurement mode.
See Figure on right.
Figure 25 – P3.2: Select number
of pH calibration points
34