Instruction Manual PD 300
16 ADDENDUM 3: METER FACTORY DEFAULT SETTINGS
Resetting the meter to factory default settings clears all calibration data and most other setup
functions you might have changed. The following settings will remain as you have set them:
• Temperature unit of measure (°C or °F)
• The temperature offset calibration value.
NOTE
DO and pH data are cleared separately from each other.
• To clear pH data, see Sub group P4.0 on page 36.
• To clear DO data, see Sub group P7.0 on page 49.
Type Parameter P No. Default Remarks
pH parameters
CAL View pH calibration data P1.1 –– Calibration data for 1st buffer, pH 1.68
P1.2 –– Calibration data for 2nd buffer, pH 4.01
P1.3 –– Calibration data for 3rd buffer, pH 7.00
P1.4 –– Calibration data for 4th buffer, pH 10.01
P1.5 –– Calibration data for 5th buffer, pH 12.45
ELE View electrode offset P2.1 0.00 mV No offset adjustment
View electrode slope P2.2 100.0 % No slope adjustment
COF Ready indicator P3.1 Ready On Ready indicator on; auto endpoint off
# pH calibration points P3.2 3 Selects number of calibration points.
Minimum 2 points.
°C or °F P3.3 No default °C or °F remains as selected
rSt pH factory default P4.0 No Retains your current settings
DO parameter
dPr Barometric pressure inputs. P1.1 Hg Hg or Pa.
760 mm Hg Adjustable from 500 – 1499 mm Hg
Salinity adjustment P1.2 0 ppt Adjust from 0 – 50.00 ppt (% only)
OFS Set % saturation offset P2.0 –– Up to ±10% (% only)
CAL View calibration data P3.0 [P2.0] –– ––
ELE View probe slope P4.1 [P3.1] 1.000 ––
View % saturation offset P4.2 [P3.2] 0% offset ––
View mV = 100% saturation P4.3 [P3.3] 37.0 mV
View mV = 0% saturation P4.4 [P3.4] 0.3 mV
COF Ready indicator P5.1 [4.1] Ready On Ready indicator on; auto endpoint off
Select mg/L or ppm units [P4.2] mg/L mg/L or ppm (mg/L mode only)
°C or °F P5.2 [P4.3] Retains settings °C or °F remains as selected
LCd Adjust back lit display P6.0 [P5.0] 8 Levels 0 to 8 (brightest)
rSt Reset to factory default P7.0 [P6.0] No Yes / no
* Program numbers with “[ ]” are accessed from mg/L (ppm) mode.