
Instruction Manual PCD 650
96
4.11 O
2
mg/L (ppm) – DO Concentration Setup
DO Concentration
(mg/L) setup screen present many options to control the operating
parameters, which can be controlled and set from the DO setup screen. The settings
are displayed in 2 pages. Press NEXT-P (F2) and PREV-P (F1) to navigate through
these pages.
4.11.1 DO (mg/L) – Page 1
Figure 88 : DO (mg/L) – Page 1
This page allows you to set operating parameters for the DO Concentration
measurement mode:
Parameter Function Factory
Default
Select DO Unit
Sets the unit of measurement for DO
Available units: mg/L and ppm
mg/L
Auto Salinity Comp
DISABLE – No automatic compensate
ENABLE – The meter will apply measured salinity for DO
concentration of your sample.
Note: This option is activated only in CD650 & PCD650.
DISABLE
Set Salinity
Allows to set the salinity (in ppt) for your solution
Available range: 0 to 50 ppt
0.0 ppt
NOTE: Auto salinity compensation will be applicable only in multi measurement
mode and only if one of the measurement selected is conductivity related. Otherwise,
manually set salinity value will be applied.
Komentáře k této Příručce