3.4 mV Calibration
Use a standard ORP solution of known value for calibration. You need to calibrate
only 1-point.
1. Switch on the meter. Make sure the meter is in mV measurement mode.
2. Press CAL (F2) to go to calibration mode.
3. The meter shows ‘Calibration-Rinse Electrode’ screen [Figure 41 (a)] few
seconds and then shows the mV calibration screen [Figure 45 (a)].
(a) (b)
Figure 45 : mV calibration screen & calibration report
4. The screen shows two readings. The upper display shows the mV reading
of the solution with respect to previous calibration (if any) & lower displays
shows the absolute mV reading of the solution without any calibration
(default reading). Use & keys to adjust the upper display to the mV
value of the solution.
5. Press ENTER key to confirm the entered value.
6. The calibration is completed. The meter shows the calibration report
[(Figure 45-(b)].
7. Press ESC (F4) to go to measurement mode. The meter shows the relative
mV reading of the solution in measurement mode, if offset is not zero.
Note: The meter allows you to adjust the upper display reading up to
±150 mV with respect to the default reading.
Note: If the meter is password protected, you will be prompted to enter a
password. Refer ‘Accessing Calibration mode when password protection
enabled’ in pa
e 42.