Instruction Manual PCD 650
43
Figure 39 : Login password screen
3. Press up & down arrow keys to enter the first digit of the password
and then press NEXT (F3) key to move to the next digit.
4. The next digit is selected. Press up & down arrow keys to enter the
second digit of the password.
5. Similarly enter all 5-digits.
6. Press ENTER key to confirm the password.
7. When the correct password is entered, the ‘Calibration – Rinse Electrode’
screen appears (Refer Figure 48-(a)].
Note: You can enter ‘00000’ (read-only password) if you wish to view the
calibration report of the last calibration. You are not allowed to perform
calibration when
y
ou enter ‘read-onl
y
password’.
Note: If you enter an incorrect password, the screen shows ‘Try again’. If
an incorrect password is entered for 3 consecutive times, the meter goes
to measurement mode.