Instruction Manual PCD 650
63
3.9 TDS Calibration
You can commence calibration in the TDS mode by using the conductivity
calibration method and after setting the correct TDS factor. For more information
regarding TDS conversion factor determination, please refer to page 106. For the
rest of the calibration process, follow steps as mentioned in the section of manual
calibration on page 55. Use a standard solution of known value for calibration. Use
the single-point standardization option to measure samples that close in value and
multi-point standardization to measure wide range of samples.
1. Switch on the meter. Make sure the meter is in TDS measurement mode.
2. Press CAL (F2) to go to calibration mode.
3. The meter shows ‘Calibration-Rinse Electrode’ screen [Figure 48-(a)] few
seconds and then shows the cell constant adjustment screen for TDS [Figure
56(a)].
(a) (b)
Figure 56 : TDS Cell constant adjustment & TDS calibration screen
4. The upper display of the TDS cell constant screen [Figure 56-(a)] shows the
TDS cell constant of your solution & lower display shows the measured value
of the solution with adjusted cell constant. Press
& keys to adjust the cell
constant for your custom solution in upper display.
5. Press ENTER key to confirm the entered value and to go to TDS calibration
screen.
6. The screen shows two readings [Figure 56-(b)]. The upper display shows the
TDS reading of the solution with respect to previous calibration (if any) & lower
display shows the TDS reading of the solution without any calibration. Use
&
keys to adjust the upper display of the TDS value of your custom solution.
7. Press ENTER key to confirm the entered value.
Note: If the meter is password protected, you will be prompted to enter a
password. Refer ‘Accessing Calibration mode when password protection
enabled’ in pa
e 42.