2.14 Working with Memory functions
The PCD 650 has a memory capacity to store up to 500 sets of measurement data.
The MEM (F2) function key allows you to view stored data. Optionally, you can
transfer this data to a Computer using wireless Infrared connection.
2.14.1 Logging data automatically in meter’s memory
1. Make sure that the Print mode is set to data logging mode in the System
Setup and you can also specify the time interval in between each transfer.
(Refer: ‘System Settings – Page 4’ on page 79)
2. From measurement mode, press PRIN (F3) to start saving data
automatically in meter’s memory. The memory location of the transferred
reading is shown in the bottom-left of the screen. (Figure 35)
Figure 35 : Saving a reading
2.14.2 Storing currently displayed reading in the memory
(In IrDA and LED print
mode)
1. Make sure you are in the measurement mode.
2. Press left or right arrow key to navigate to other available functions
until you see STOR function in the LCD.
3. Press STOR (F3) key to store the currently displayed reading. The display
briefly shows the memory location where this reading is being saved.
(Figure 35)
2.14.3 Viewing stored data
1. Make sure you are in measurement mode.
2. Press left or right arrow key to navigate to other available functions
until you see MEM function in the LCD.
Memory
Location
where
readings
are being
saved.
NOTE: This
symbol will
appear if data
logging mode
has been
selected in
“System set
up’. It will
automatically
start saving
data in meter’s
memory, when
PRIN (F3) key
is pressed.